Functional Principle Myo Electrics
Gripping force that adjusts itself

Digital technology for the hand: DMC plus can be used with residual limbs of almost any length, except wrist disarticulation residual limbs. In DMC plus control mode, a higher signal is required to open the hand after grasping once with maximum gripping force. This reduces the risk of opening the hand with unintended muscle signals.

  • Two independent measurement and control systems proportionally control both the gripping speed and gripping force.
  • The gripping speed and gripping force are determined by the strength of the muscle signal.
  • The system electric hand DMC plus can be operated with the MyoEnergy Integral, the EnergyPack or the X-ChangePack.

Product information

Technical data
Operating voltage 6/7.2 V
Opening width Size 7 [79 mm], sizes 7 ¼, 7 ¾, 8 ¼ [100 mm]
Weight [with system inner hand] Size 7 [355 g], sizes 7 ¼, 7 ¾, 8 ¼ [457 g]
Sizes 7, 7 ¼, 7 ¾, 8 ¼
Proportional speed 15–130 mm/sec
Proportional gripping force 0–90 N
Functional Principle Myo Electrics